LSM 510 OPERATION IN EXPERT MODELSM 510 META Macro Menu Carl Zeiss03/06 B 45-0021 e 4-183To match the scan fields perform the followingsteps:• Use the specific slide provided and focus ontothe grid structure.• Start the ScanfieldTransform Macro.• Chose a scanner by ticking the correspondingcheck box.• Tick the checkbox for the Test grid.• Set up a track for reflection imaging with thescanner chosen in the Macro.• Start a continuous scan and position the center of the grid in the center of the image window.• If necessary refocus onto the test slide. The projection of the Test grid into the image can be selectedand deselected while scanning.• Position the test slide such that the center point position matches the center point position of the Testgrid projection.• Use the sliders for Offset, Amplitude and Rotation to adjust the scanned image to the projected Testgrid.Fig. 4-177 Sample Configuration for imaging thetest slide in reflection using the ArgonLaser of the LSM 510Fig. 4-178 Projection of the Test grid into the image window