CHAPTER 4 - SYSTEM OPERATIONLightsheet Z.1 Left Tool Area and Hardware Control Tools Carl Zeiss02/2013 000000-1790-528 753.2.11.4 EndExpand the End field by clicking the triangle(Fig. 114).In the Mode drop down menu (Fig. 115),you can select the way the acquisition istriggered. You have the choice betweenManual and Time.If Manual is selected the acquisition of atime series ends using the Stop button orafter the defined number of cycles.If Time is selected the hh (hour), min(minute) and sec (second) input boxesbecome available (Fig. 116), in which thehours, minutes and seconds can be edited.The system interrupts the acquisition afterthe defined computer time has beenreached.A Time Series is finished when the end Time is reached.If the entered number of Cycles has been processed, the Time Series is finished before the endTime. If the number of cycles has not yet been processed until the end Time, the Time Series is alsofinished. The end time for the Time Series can be changed online.If a time series is interrupted before its programmed end, the programmed number of images willbe taken over in the database. However, only those images are stored which were created beforeinterruption of the time series. This is due to the fact that the original image parameters are to betaken over via the Reuse function.Fig. 114 End fieldFig. 115 End Mode drop down menuFig. 116 End Mode drop down menu – Time selected