CHAPTER 1 - HARDWARELightsheet Z.1 User Interfaces Carl Zeiss02/2013 000000-1790-528 273.3.2 Assembly of the Sample Chamber Body and the Sample Chamber Dove Tail Slide− Place the mesh screen (Fig. 20/2) into the middle of the sample chamber body (Fig. 20/1) to reduceturbulence stress on the specimen when the chamber is filled.− For cleaning, it can be carefully removed by a forceps or via a teflon-tip-free end of a capillaryplunger.− If a Heating component is used, proceed first with chapter 3.7.1 (see PeCon manual for specificheating component) and screw it onto the bottom of sample chamber body.• The sample chamber body (Fig. 21/3) can now be attached to the sample chamber dove tail slide(Fig. 21/1).• Lay down the sample chamber dove tail slide with the two positioning pins facing upwards (Fig. 21/2).• Take the sample chamber body and position it on to the sample chamber dove tail slide with thedetection optic adapter window facing to the back relative to the sample chamber dove tail slide grip.• Invert the sample chamber body with the sample chamber dove tail slide so that the four screw-holes(Fig. 21/4) that were on the bottom are now facing you.• Insert the four screws (diameter = M3, length = 14 mm) (Fig. 21/5) into the four screw-holes andtighten using and Allen wrench (2.5).1 Mesh screen2 Sample chamber bodyFig. 20 Placing the mesh screen