CHAPTER 1 - HARDWARECarl Zeiss User Interfaces Lightsheet Z.122 000000-1790-528 02/20133.2.4 Cable Connections for the Detection Module "PCO.Edge"The detection module "PCO.Edge" is equipped with a liquid cooling system. Please read the safety notesof the Lightsheet Z.1 system, as well as the separate safety data sheet for the liquid cooling system. Thepower supply unit (PSU) for the cooling is connected to the system's power strip at the rear of the lasermodule. The cooling hoses should be as short as possible. The connection creates a closed circuit, startingfrom the cooling unit leading to the detection module 1 (approx. 200 cm hose length) reflection port(Cam1). From the latter a hose (approx. 200 cm) leads either back to the cooling unit or, if a seconddetection module is used, first to the transmission port (Cam2) (approx. 100 cm) and then a third hose(approx. 200 cm) leads back to the cooling unit.The detection module cables are connected as follows:− CameraLink cables A and B are each plugged into the PC for system control. The upper slots of thePC for system control are assigned to the detection module of the reflection port (Cam1), [CamLinkA] right and [CamLink B] left. The lower slots are assigned to the transmission port (Cam2),[CamLink A[ right and [CamLink B] left.− The power cable is connected to the power strip of the system on the rear of the LB RackLightsheet.− The trigger cable is always connected to the detection module with [1/IN] and on the rear side ofthe main system module Lightsheet Z.1 to [Camera 1/IN] (refection port, Cam1) or [Camera 2/IN](transmission port, Cam2).− The detection module is switched on with the toggle switch on the module itself.Fig. 16 Cable connections for the detection module "PCO.Edge"