CHAPTER 1 - HARDWARELightsheet Z.1 User Interfaces Carl Zeiss02/2013 000000-1790-528 473.10 Removing and Inserting the Reflector Turret for Emission SelectionThe reflector turret (Fig. 37/5) can be accessed by opening the left-hand-side front system door(Fig. 37/1).After the reflector turret for emission selection has been exchanged, the Automatic DetectorAlignment should be performed. The ZEN software should be restarted if the reflector turret wasremoved, while ZEN was still running.To remove, first of all turn the silver-colored lever (Fig. 37/2), then pull-out the reflector turret by usingthe grip (Fig. 37/6).When inserting, it is recommended that the reflector turret first of all be held by the short grip (Fig. 37/5)and inserted into the guide rail (Fig. 37/3) and then pushed in as far as it will go using the long pull rod(Fig. 37/6). Return the locking lever to its original position. Finally, close the system door.The filters are not push and click filters and cannot be taken out as a whole. The filter cube and thebeam splitter were installed and aligned in the factory. Do not try to take out single filter cubes orbeam splitters, it will damage the filters and the alignment, rendering the system in the worst caseunusable.1 Front system door 4 Reflector turret2 Silver-colored lever 5 Short grip3 Guide rail 6 GripFig. 37 Removing and inserting the reflector turret for emission selection