CHAPTER 1 - HARDWARELightsheet Z.1 User Interfaces Carl Zeiss02/2013 000000-1790-528 313.4 Removing and Inserting the Sample ChamberThe sample chamber can be removed from the system. For removal, proceed as follows:− Open the front system cavity door (Fig. 25/1) of the main system module Lightsheet Z.1.− Remove the liquid from the sample chamber (Fig. 25/6) using the hose with the correspondingsyringe (Fig. 25/9).− Ensure that no sample protrudes into the sample chamber. Remove the sample, if necessary,through the upper sample opening (Fig. 25/2) below the upper system cavity door (Fig. 25/3).− If necessary, remove the connections for the incubation device (Fig. 25/8).− Loosen the screw (Fig. 25/4) on the sample chamber and pull the sample chamber out by the grip(Fig. 25/5).− Inserting the sample chamber:− Take hold of the sample chamber by the lower grip (Fig. 25/5) and push the sample chamber(Fig. 25/6) into the guide rails (Fig. 25/7).− Tighten the screw (Fig. 25/4) without exerting force. The sample chamber has been correctlyinserted when its lower edge is flush with the guide rail.− Plug in the cable to the incubation unit (Fig. 25/8) as necessary.1 Cover for capillaries (3 mm opening)2 Cover for syringes (7 mm opening)3 Sample chamber body with dove tail slideFig. 24 Placing cover on top of the sample chamber body