CHAPTER 1 - HARDWARECarl Zeiss User Interfaces Lightsheet Z.130 000000-1790-528 02/20133.3.4 Insertion of Accessories for Incubation• A temperature sensor (Fig. 23/1) can be used in the sample chamber body.− Hold the temperature sensor so that it shows you the letter “J”. Slide the temperature sensorsleeve (Fig. 23/3) onto the top horizontal part of the J-shaped temperature sensor.− Dip the temperature sensor with sleeve into the sample chamber body and position thetemperature sensor sleeve securely into the sample chamber body groove (Fig. 23/2).− Place a cover on top of the sample chamber body (Fig. 24/3). This will minimize evaporation ofmedium or contamination of the sample chamber. If CO 2 module Lightsheet Z.1 is used(configuration dependent, chapter 2.7.2) the cover is recommended to maintain the CO2 gasconcentrations.− Keep the opening as small as possible. Use the 3 mm opening cover for capillaries (Fig. 24/1) and7 mm opening cover for syringes (Fig. 24/2). The cover only rests on top of the sample chamberbody and is free to move.1 Temperature sensor2 Sample chamber body groove3 Temperature sensor sleeveFig. 23 Insertion of temperature sensor