CHAPTER 4 - SYSTEM OPERATIONCarl Zeiss Left Tool Area and Hardware Control Tools Lightsheet Z.1172 000000-1790-528 02/20133.4.5 Maintain – AdjustmentIn the Adjustment tool window, there are threebuttons that start tools for the detection modulerecognition and alignment (Fig. 269).Detect Camera Order, Automatic DetectorAlignment, and Manual Detector Alignmentare inherently part of the detection moduleexchange procedure.When you wish to exchange detection modules, please follow the complete descriptions inCHAPTER 1 HARDWARE.Using the show all mode for this tool window, a fourth button appears: Initial Light Sheet Alignment.This is a task for authorized service personnel of Carl Zeiss Microscopy GmbH only and is thereforepassword protected.Do not use the Initial Light Sheet Alignment service task. You run the risk of altering elementarysettings for your Lightsheet Z.1 system that have been optimally set upon delivery and installation.Please contact your local Carl Zeiss representative regarding this point only if you have designedand built your own sample chamber.3.4.5.1 Adjustment – Detector RecognitionWhen the detection modules are exchanged or attached to the Lightsheet Z.1 system for the first time,the system must identify which detection module is present for which channel.Pressing the button Detector recognition will start an identification routine to do this job. A progressbar is visible during the process.When the Cancel button is pressed, the operation is aborted without result.This routine must be conducted before the Automatic or Manual Detector Alignment.At the end of this process, you will be informed if the detector recognition had to perform any changes.If so, please restart ZEN software before continuing.Fig. 269 Adjustment tool window